Technical parameters |
||||||||||||||||
Model |
GX30B |
GX30B |
GX45B |
GX125B |
GX230B |
|||||||||||
GX30BE |
GX30BE |
GX45BE |
GX125BE |
GX230BE |
||||||||||||
Application |
high efficient physical sterilizer with easy operation |
|||||||||||||||
Air circulation type |
double air flues and forced convection |
|||||||||||||||
Performance |
temp. range |
RT+10~300℃ |
||||||||||||||
temp. resolution |
0.1℃ |
|||||||||||||||
temp. fluctuation |
±1℃ |
|||||||||||||||
temp. uniformity |
±2.5% |
|||||||||||||||
Structure |
chamber material |
stainless steel |
||||||||||||||
shell material |
spraying cold rolled steel |
|||||||||||||||
thermal insulation material |
aluminum silicate fiber |
|||||||||||||||
heating device |
stainless heating tube |
nichrome heating wire |
||||||||||||||
rated power |
0.8kw |
1.2kw |
1.6kw |
2.3kw |
3.0kw |
|||||||||||
vent |
φ28mm*1 (top) |
φ28mm*2 (top) |
||||||||||||||
Controller |
temp. control |
model B: double rows of 4 digital tube, PID; |
||||||||||||||
temp. setting |
press the four keys slightly |
|||||||||||||||
temp. display |
model B: actual temperature is displayed by 4 digit digital tube (row 1); |
|||||||||||||||
model BE: actual temperature. is displayed by LCD (row 1); |
||||||||||||||||
timer |
0~9999minutes(with timing function) |
|||||||||||||||
operation function |
running with data set and time set, and automatic shut-down based on setting |
|||||||||||||||
program mode |
optional |
|||||||||||||||
sensor |
Pt100 |
|||||||||||||||
additional function |
deviation calibration, menu buttons lock, power outage compensation, power outage memory |
|||||||||||||||
Safety device |
model B and BE: over-temperature alarm; |
|||||||||||||||
Specification |
working size |
310*310*310 |
350*350*350 |
400*360*450 |
500*450*550 |
600*500*750 |
||||||||||
exterior size |
460*510*695 |
500*550*735 |
550*550*840 |
636*680*915 |
730*670*1220 |
|||||||||||
packing size |
550*570*765 |
620*585*800 |
640*640*905 |
730*720*1000 |
900*800*1350 |
|||||||||||
working capacity |
30L |
45L |
65L |
125L |
230L |
|||||||||||
shelf load bearing |
15kg |
|||||||||||||||
max shelf layers |
6layers |
7layers |
9layers |
13layers |
17layers |
|||||||||||
height between shelves |
35mm |
|||||||||||||||
Power source( 50/60Hz)/Rated current |
AC220V/3.6A |
AC220V/5.5A |
AC220V/7.2A |
AC220V/10.5A |
AC220V/13.6A |
|||||||||||
NW/GW |
33/37kg |
37/43kg |
44/49kg |
60/66kg |
94/120kg |
|||||||||||
Accessory |
shelf |
2pcs |
||||||||||||||
shelf bracket |
4pcs |
|||||||||||||||
Optional device |
shelf, RS485 port, printer, recorder, external communication, remote control, program temperature controller, wireless message alert, U disk storage. |
|||||||||||||||
Remarks |
fan of GX230B and GX230BE model is fixed speed type |
|||||||||||||||
update the heating device to far-infrared heating device, it can be far-infrared drying oven and save 40% energy consumption. |