Features:
? Classic benchtop pH meter is equipped with a white backlit LCD display.
? 2 points push-button calibration with auto-buffer recognition.
? Electrode slope display helps user decide whether to replace pH electrode.
? Manual temperature compensation improves the measurement accuracy of meter.
Applications:
Education: chemical experiments
Laboratory: routine measurements and quality controls
Specifications
Model |
PHS-3CW |
|
pH |
||
Range |
0.00~14.00pH |
|
Accuracy |
±0.01pH |
|
Resolution |
0.01pH |
|
Calibration Points |
2 points |
|
Calibration Solutions |
USA (pH4.01/7.00/10.01) or NIST (pH4.01/6.86/9.18) |
|
Temperature Compensation |
0~100°C |
|
Compensation Mode |
Manual |
|
mV |
||
Range |
-1999~1999mV |
|
Accuracy |
±1mV |
|
Resolution |
1mV |
|
Others |
||
Connector |
BNC |
|
Power Requirements |
DC9V, using AC adapter, 220VAC/50Hz |
|
Dimensions |
210(L)×205(W)×65(H)mm |
|
Weight |
1.5kg |
Order Information
Order Code |
Accessories |
PHS-3CW-CN |
E201-BNC pH Electrode |
pH Buffer Pouches (pH4.01/7.00/10.01) |
|
DC9V Power Adapter |
|
PHS-3CW-UK
|
P11 Glass pH Electrode (Made in UK) |
pH Buffer Pouches (pH4.01/7.00/10.01) |
|
DC9V Power Adapter |
Optional Accessories
Items |
Description |
Suitable for measuring the pure water |
|
Suitable for measuring the general liquids |
|
Suitable for measuring the semisolid substances |
|
pH4.01/7.00/10.01 |
|
pH4.01/6.86/9.18 |
Packaging
Order Code |
Dimensions |
Weight |
PHS-3CW-CN |
310(L)×240(W)×180(H)mm |
1.7kg |
PHS-3CW-UK |
310(L)×240(W)×180(H)mm |
1.7kg |