8-channel Adjustable Volume Pipettes |
||||||
Volume Range |
Increment |
Test Volume |
Maximum permissible systematic error (Inaccuracy) |
Maximum permissible random error (Imprecision) |
||
% |
μL |
% |
μL |
|||
0.5-10μL |
0.1μL |
10μL |
1.50% |
0.15 |
1.50% |
0.15 |
5μL |
2.50% |
0.125 |
2.50% |
0.125 |
||
1μL |
4.00% |
0.04 |
4.00% |
0.04 |
||
5-50μL |
0.5μL |
50μL |
1.00% |
0.5 |
0.50% |
0.25 |
25μL |
1.50% |
0.375 |
1.00% |
0.25 |
||
5μL |
3.00% |
0.15 |
2.00% |
0.1 |
||
50-300μL |
5μL |
300μL |
0.70% |
2.1 |
0.25% |
0.75 |
150μL |
1.00% |
1.5 |
0.50% |
0.75 |
||
50μL |
1.50% |
0.75 |
0.80% |
0.4 |
||
12-channel Adjustable Volume Pipettes |
||||||
Volume Range |
Increment |
Test Volume |
Maximum permissible systematic error (Inaccuracy) |
Maximum permissible random error (Imprecision) |
||
% |
μL |
% |
μL |
|||
0.5-10μL |
0.1μL |
10μL |
1.50% |
0.15 |
1.50% |
0.15 |
5μL |
2.50% |
0.125 |
2.50% |
0.125 |
||
1μL |
4.00% |
0.04 |
4.00% |
0.04 |
||
5-50μL |
0.5μL |
50μL |
1.00% |
0.5 |
0.50% |
0.25 |
25μL |
1.50% |
0.375 |
1.00% |
0.25 |
||
5μL |
3.00% |
0.15 |
2.00% |
0.1 |
||
50-300μL |
5μL |
300μL |
0.70% |
2.1 |
0.25% |
0.75 |
150μL |
1.00% |
1.5 |
0.50% |
0.75 |
||
50μL |
1.50% |
0.75 |
0.80% |
0.4 |
Fixed Volume Pipettes |
||||||
Volume Range |
Increment |
Test Volume |
Maximum permissible systematic error (Inaccuracy) |
Maximum permissible random error (Imprecision) |
||
% |
μL |
% |
μL |
|||
5μL |
- |
5μL |
1.3% |
0.065 |
1.2% |
0.06 |
10μL |
- |
10μL |
0.8% |
0.08 |
0.8% |
0.08 |
20μL |
- |
20μL |
0.6% |
0.12 |
0.5% |
0.1 |
25μL |
- |
25μL |
0.5% |
0.125 |
0.3% |
0.075 |
50μL |
- |
50μL |
0.5% |
0.25 |
0.3% |
0.15 |
100μL |
- |
100μL |
0.5% |
0.5 |
0.3% |
0.3 |
200μL |
- |
200μL |
0.4% |
0.8 |
0.2% |
0.4 |
250μL |
-- |
250μL |
0.4% |
1.0 |
0.2% |
0.5 |
500μL |
- |
500μL |
0.3% |
1.5 |
0.2% |
1.0 |
1000μL |
- |
1000μL |
0.3% |
3.0 |
0.2% |
2.0 |
2000μL |
- |
2000μL |
0.3% |
6.0 |
0.15% |
3.0 |
5000μL |
- |
5000μL |
0.3% |
15 |
0.15% |
7.5 |