1.Light Source:Hamamatsu 150W Xenon Lamp
2. Exciting Optical Filters:Interference Optical Filter
*Standard Set Is Equipped With An Interference Optical Filter Of Central Wavelength At 365Nm And 10Nm Bandwidth
*An Optional Interference Optical Filer Of 25Mm Diameter Of Wavelength Of 250-700 Nm Can Be Provided On Users Demand.
3.Emission Monochromator:C-T Diffraction Grating(Em 200~900Nm,Bandwidth 10Nm)Wavelength Accuracy±1Nm Wavelength Repeatability ≤0.5Nm
4.Sensitivity:Raman Peak Of Water In 1 Cm Quartz Fluorescence Cuvette With
S/N≥150
5.Linear Measurement(R):≥0.995
6.Stability: Better Than 1.5%/10Min
7.Variation Of Power Source: 220V±22V 50Hz±1Hz
8.Response Time: (0.1-4)S 6 Stages Adjustable
9.Fluorescence Display Value: 0.00-600.00
10. Data Transmission: Usb2.0
11.Software:Including